GHSA-9p7r-jmp9-7hw2MediumCVSS 4.0
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling...
🔗 CVE IDs covered (1)
📋 Description
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.